Bug log

Findings awaiting review. Raised 2026-08-21 during the R11 source-derived scorer vector work.

Each B-* finding is pinned by a characterisation test in test/scoring/scorer-source-vectors.test.js. Those tests assert current behaviour so that any change is deliberate and visible. They are explicitly not an endorsement that the current behaviour is clinically correct.

Nothing in this log has been clinically reviewed. B-01, B-02, B-05 and B-06 change numeric PEWS output if resolved, so they are clinical-behaviour decisions, not refactors.

Scorer findings

All six share one root cause: scoreFromBands and scoreOxygen in chart/npews-scorer.js return 0 when no rule matches. A score of 0 is also the legitimate value for a normal observation, so an unmatched, malformed or unrecorded input is indistinguishable from a normal one. Every failure below therefore resolves towards falsely reassuring.

ID Finding Current behaviour Why it matters
B-01 The canonical bands are not contiguous. Upper bounds use .99 style values ([0, 9.99, 4] then [10, 19.99, 2]), leaving the open interval (9.99, 10) undeclared. Scores 0. Unreachable for integer-valued vitals, but reachable for any parameter a device reports with more than two decimal places. Both neighbouring bands may score above 0.
B-02 A value beyond the outermost declared bounds matches no band. Scores 0. oxygenDeliveryLpm tops out at [6, 20, 4], so a level device charted at 25 L/min scores 0 rather than 4. A unit-confusion or transcription error reads as "normal".
B-03 scoreObservation performs no numeric validation, unlike deriveNonScoreEscalation which throws on unknown avpu, newSepsisSuspicion and non-finite temperature. NaN and Infinity score 0 silently. NaN fails every >= / <= comparison, so a corrupt reading is silently absorbed into the total.
B-04 The canonical spec records that capillary refill 2.01-2.99s has no defined score. scoreCrt resolves the hole downwards to 0. A known specification hole is currently closed by implementation accident rather than by decision.
B-05 scoreOxygen requires LEVEL_DEVICES.has(device) && delivery. A level device with oxygenDelivery: null scores 0. A patient demonstrably receiving oxygen scores identically to one breathing air. Relates to H-006 in SAFETY.md.
B-06 An unrecognised oxygenDevice code is neither high-flow nor level. Scores 0 silently, no error. Inconsistent with the strict validation applied to other coded fields. A host mapping defect fails silently.

Options for resolution

These interact, so they should be decided together rather than piecemeal:

  1. Clamp to the nearest declared band. Extreme values score as the most abnormal band. Keeps pewsTotal a plain number; masks genuine data errors.
  2. Throw on unmatched input. Consistent with existing validation. Requires every host to supply clean data, and an uncaught throw in a rendering path is its own hazard.
  3. Return an explicit unscoreable marker. Safest clinically, but changes the contract of pewsTotal / scoreBreakdown and therefore affects the chart, the FHIR adapter, escalation selection and SAFETY.md.

Option 3 overlaps substantially with R53 ("make the missing BP explicit with its applicable reason") and with the validation contract already queued as SC-011. Recommend deciding it once, there, rather than separately here.

Documentation discrepancies

ID Finding
D-01 roadmap.md records Last reviewed: 2026-08-06, but the uncommitted changes incorporate clinical review through 2026-08-11 and further review on 2026-08-21.
D-02 The roadmap section heading Doing Next - Settle The Chart UI no longer matches the stated priority. The intro now names requirements traceability and fixtures as immediate, and R54 has moved to Stage 5. Every item in the section is [x] except R41 and R50.
D-03 R50 sits under "Doing Next" but is blocked externally on NHSE supplying authoritative RGB values (issue #5). It is not actionable by this team.
D-04 requirements-traceability.md marks C8.6 and C15.2 as Partial, which is not one of the five allocation states its own legend defines.
D-05 C1.1 is allocated twice in the traceability matrix, to two different buckets (observation entry/workflow, and interoperability/coding).
D-06 The C8.2 evidence cell reads C17.11 below rather than citing code or a test, unlike every other row.

Withdrawn

ID Finding Outcome
W-01 Suspected that the U3.5, U3.6 and U3.11 "available by interaction" claims in the traceability matrix were unsubstantiated. Not a defect. attachTooltip and attachCategoricalTooltip implement hover tooltips, chart.js:11 cites U3.5/U3.6 directly, and canvases carry aria-label text alternatives. Worth noting only that the tooltips are mousemove-driven with no keyboard/focus equivalent, so keyboard-only users depend on the aria-label — a point for R14, not a traceability error.